Agilent Technologies 20ET ユーザーズマニュアル

ページ / 478
6-36
 
Calibrating for Increased Measurement Accuracy
Power Meter Measurement Calibration
NOTE
Remember to subtract the through arm loss from the coupler arm loss before 
entering it into the power loss table, to ensure the correct power at the output 
of the coupler.
4. Repeat the previous two steps to enter up to 12 frequency segments, depending on the 
required accuracy.
You may enter multiple segments in any order because the analyzer automatically sorts 
them and lists them on the display in increasing order of frequency.
If you only enter one frequency segment, the analyzer assumes that the single value is 
valid over the entire frequency range of the correction.
5. After you have entered all the segments, press 
.
6. Press 
 
 
 to activate the power loss compensation.
Using Sample-and-Sweep Correction Mode 
You can use the sample-and-sweep mode to correct the analyzer output power and update 
the power meter correction data table, during the initial measurement sweep. Because the 
analyzer measures the actual power at each frequency point during the initial sweep, the 
initial sweep time is significant. However, in this mode of operation the analyzer does not 
require the power meter for subsequent sweeps. Therefore, this mode sweeps considerably 
faster than the continuous correction mode.
Figure 6-10 Sample-and-Sweep Mode for Power Meter Calibration
1. Calibrate and zero the power meter.
2. Connect the equipment as shown in 
.
 DONE 
Cal
 PWRMTR CAL   PWR LOSS ON