Agilent Technologies 20ET ユーザーズマニュアル

ページ / 478
6-50
 
Calibrating for Increased Measurement Accuracy
Calibrating for Non-Coaxial Devices (ES Analyzers Only)
Calibrating for Non-Coaxial Devices (ES Analyzers Only)
The analyzer has the capability of making calibrations using the TRL*/LRM* method. 
TRL* and LRM* are implementations of the thru-reflect-line and line-reflect-match 
calibrations, modified for the three-sampler receiver architecture in the analyzer. 
TRL/LRM calibration is available in analyzers with Option 400, which adds a fourth 
sampler to the receiver. The following procedures apply to both TRL*/LRM* and TRL/LRM 
calibrations.
TRL Error Correction 
Create a User-Defined TRL Calibration Kit 
In order to use the TRL technique, the calibration standards characteristics must be 
entered into the analyzer’s user defined calibration kit. 
The following steps show you how to define a calibration kit to utilize a set of TRL (THRU, 
REFLECT, LINE) standards. This example TRL kit contains the following:
• zero length THRU
• "flush" short for the REFLECT standard (0 second offset)
• 50 ohm transmission line with 80 ps of offset delay for the LINE
Modify the Standard Definitions
1. Press the following keys to start modifying the standard definitions:
 
 
 
 
2. To select a short, press 
 
. (In this example, the 
 standard is a 
.)
3. Press the following keys:
 
 
 
 
 
 
 
4. To define the THRU/LINE standard, press:
 
 
 
 
 
 
 
 
 
 
5. To define the LINE/MATCH standard, press:
 
 
 
 
 
 
 
 
 
 
Enter a frequency greater than the maximum frequency range of the analyzer. For 
example, press 
 
. Then, press 
.
Preset
Cal
 CAL KIT   MODIFY   DEFINE STANDARD 
1
x1
 REFLECT 
 SHORT 
 SHORT   MODIFY STD DEFINITION  SPECIFY OFFSET   OFFSET DELAY
0
x1
 STD OFFSET DONE   STD DONE (DEFINED) 
 DEFINE STANDARD 
4
x1
 DELAY/THRU   MODIFY STD DEFINITION 
 SPECIFY OFFSET   OFFSET DELAY 
0
x1
 STD OFFSET DONE 
 STD DONE (DEFINED) 
 DEFINE STANDARD 
6
x1
 DELAY/THRU   MODIFY STD DEFINITION
 SPECIFY OFFSET   OFFSET DELAY 
.08
G/n
 MAXIMUM FREQUENCY
50
G/n
 STD OFFSET DONE