Calibrating for Increased Measurement Accuracy
Adapter Removal Using ECal (ES Analyzers Only)
Adapter Removal Using ECal (ES Analyzers Only)
A device under test (DUT) whose connectors cannot be connected directly to a test 
configuration is considered to be a noninsertable device. See 
 Noninsertable 
devices can be caused because the DUT has:
• Input or output connectors with the same sex connector as the test configuration.
• Input or output connectors of a different connector type than the test configuration.
Figure 6-25 Noninsertable Device
The adapter removal calibration technique may be used with the ECal calibration to 
accurately measure noninsertable devices. The following adapters (shown in 
are needed.
• Adapter A1, which mates with port 1 of the DUT, must be installed on Port 1 of the 
analyzer.
• Adapter A2, which mates with port 2 of the DUT, must be installed on Port 2 of the 
analyzer.
• Adapter A3 must match the connectors on the DUT. The effects of this adapter will be 
completely removed with this calibration technique.
NOTE
Adapters A1 and A2 become part of the test setup to allow connection to the 
DUT. Adapter A3 is used during the calibration only. Its effects will be 
removed.