7- 5
Operating Concepts
Processing
 Processing 
The analyzer’s receiver converts the R, A, and B input signals into useful measurement 
information. This conversion occurs in two main steps:
• The swept high frequency input signals are translated to fixed low frequency IF signals, 
using analog sampling or mixing techniques. (Refer to the service guide for more details 
on the theory of operation.)
• The IF signals are converted into digital data by an analog to digital converter (ADC). 
From this point on, all further signal processing is performed mathematically by the 
analyzer microprocessors.
The following paragraphs describe the sequence of math operations and the resulting data 
arrays as the information flows from the ADC to the display. They provide a good 
foundation for understanding most of the response functions, and the order in which they 
are performed.
 is a data processing flow diagram that represents the flow of numerical data 
from IF detection to display. The data passes through several math operations, denoted in 
the figure by single line boxes. Most of these operations can be selected and controlled with 
the front panel response block menus. The data, stored in arrays along the way and 
denoted by double line boxes, are places in the flow path where data is accessible via GPIB.
Figure 7-2 Data Processing Flow Diagram