Agilent Technologies 20ET ユーザーズマニュアル

ページ / 478
7- 11
Operating Concepts
Sweep Time
Sweep Time 
The 
 softkey selects sweep time as the active entry and shows whether 
the automatic or manual mode is active. The following explains the difference between 
automatic and manual sweep time:
• Manual sweep time. As long as the selected sweep speed is within the capability of 
the instrument, it will remain fixed, regardless of changes to other measurement 
parameters. If you change measurement parameters such that the instrument can no 
longer maintain the selected sweep time, the analyzer will change to the fastest sweep 
time possible.
• Auto sweep time. Auto sweep time continuously maintains the fastest sweep speed 
possible with the selected measurement parameters.
Sweep time refers only to the time that the instrument is sweeping and taking data, and 
does not include the time required for internal processing of the data, retrace time, or band 
switching time. A sweep speed indicator 
↑ 
is displayed on the trace for sweep times longer 
than 1.0 second. For sweep times equal to or faster than 1.0 second, the 
 indicator appears 
in the status notations area at the left of the analyzer's display.
Manual Sweep Time Mode 
When this mode is active, the softkey label reads 
. This mode is 
engaged whenever you enter a sweep time greater than zero. This mode allows you to 
select a fixed sweep time. If you change the measurement parameters such that the 
current sweep time is no longer possible, the analyzer will automatically increase to the 
next fastest sweep time possible. If the measurement parameters are changed such that a 
faster sweep time is possible, the analyzer will not alter the sweep time while in this mode.
Auto Sweep Time Mode 
When this mode is active, the softkey label reads 
. This mode is 
engaged whenever you enter 
 
 as a sweep time. Auto sweep time continuously 
maintains the fastest sweep time possible with the selected measurement parameters.
Minimum Sweep Time 
The minimum sweep time is dependent on the following measurement parameters:
• the number of points selected
• IF bandwidth
• sweep-to-sweep averaging in dual channel display mode
• error-correction
• type of sweep
SWEEP TIME [ ] 
SWEEP TIME [MANUAL] 
SWEEP TIME [AUTO] 
0
x1