Agilent Technologies 20ET ユーザーズマニュアル

ページ / 478
7- 33
Operating Concepts
Electrical Delay
Electrical Delay 
The 
 softkey adjusts the electrical delay to balance the phase of the 
test device. This softkey must be used in conjunction with 
 or 
 (with cut-off frequency) in order to identify which type of 
transmission line the delay is being added to. These softkeys can be accessed by pressing 
the  
key.
Electrical delay simulates a variable length lossless transmission line, which can be added 
to or removed from a receiver input to compensate for interconnecting cables, etc. This 
function is similar to the mechanical or analog "line stretchers" of other network analyzers. 
Delay is annotated in units of time with secondary labeling in distance for the current 
velocity factor.
With this feature, and with 
 (refer to 
), an equivalent length of air-filled, lossless transmission line is added or 
subtracted according to the following formula:
Once the linear portion of the test device's phase has been removed, the equivalent length 
of the lossless, transmission line can be read out in the active marker area. If the average 
relative permittivity (
ε
r
) of the test device is known over the frequency span, the length 
calculation can be adjusted to indicate the actual length of the test device more closely. 
This can be done by entering the relative velocity factor for the test device using the 
calibrate more menu. The relative velocity factor for a given dielectric can be calculated by:
assuming a relative permeability of 1.
ELECTRICAL DELAY 
COAXIAL DELAY 
WAVEGUIDE DELAY 
Scale Ref
MARKER 
 DELAY 
Length meters
(
)
Φ
Freq MHz
(
)
1.20083
×
(
)
-------------------------------------------------------------------
=
Velocity Factor
1
ε
r
--------
=