Operating Concepts
TRL*/LRM* Calibration (ES Models Only)
TRL Terminology 
Notice that the letters TRL, LRL, LRM, etc. are often interchanged, depending on the 
standards used. For example, "LRL" indicates that two lines and a reflect standard are 
used; "TRM" indicates that a thru, reflection and match standards are used. All of these 
refer to the same basic method.
TRL* calibration is a modified form of TRL calibration. It is adapted for a receiver with 
three samplers instead of four samplers. The TRL* calibration is not as accurate as the 
TRL calibration because it cannot isolate the source match from the load match, so it 
assumes load match and source match are equal. However, option 400 on the ES model 
adds the fourth sampler and provides for a true TRL calibration.
How TRL*/LRM* Calibration Works 
The TRL*/LRM* calibration used in the analyzer relies on the characteristic impedance of 
simple transmission lines rather than on a set of discrete impedance standards. Since 
transmission lines are relatively easy to fabricate (in a microstrip, for example), the 
impedance of these lines can be determined from the physical dimensions and substrate’s 
dielectric constant.
TRL* Error Model 
Figure 7-43 Functional Block Diagram for a 2-Port Error-Corrected 
Measurement System 
For the analyzer TRL* 2-port calibration, a total of 10 measurements are made to quantify 
eight unknowns (not including the two isolation error terms). Assume the two 
transmission leakage terms, E
XF
 and E
XR
, are measured using the conventional technique. 
The eight TRL error terms are represented by the error adapters shown in 
Although this error model is slightly different from the traditional Full 2-port 12-term 
model, the conventional error terms may be derived from it. For example, the forward 
reflection tracking (E
RF
) is represented by the product of 
ε
10
 and 
ε
01