Agilent Technologies 20ET ユーザーズマニュアル

ページ / 478
1- 59
Making Measurements
Measuring Amplifiers
Making High Power Measurements with Option 085
(ES Analyzers Only)
Analyzers equipped with Option 085 can be configured to measure high power devices. 
This ability is useful if the required input power for a device under test is greater than the 
analyzer can provide, or if the maximum output power from an amplifier under test 
exceeds safe input limits for a standard analyzer. This section describes how to set up the 
analyzer to perform high power measurements.
Initial Setup 
1. If the analyzer is in the bypass mode configuration, remove the jumper between the RF 
OUT and RF IN connector on the rear panel.
2. Connect the booster amplifier RF INPUT connector to the RF OUT connector on the 
rear panel of the analyzer.
3. Connect a 20 dB coupler (that operates within the frequency range of interest) to the 
booster amplifier RF OUTPUT connector.
Figure 1-47 High Power Test Setup (Step 1) 
Determining Power Levels 
Before continuing, save this state and rename it to UPRESET. To allow the preset mode to 
be set to user preset, press 
 
. The final state can then be 
resaved as a user preset to avoid an over-power condition from the factory preset. 
4. Switch on the analyzer and reduce the power level to 
20 dBm by pressing 
 
 
 
 
.
Preset
  PRESET: USER 
Power
 PWR RANGE AUTO   POWER RANGES 
20
x1