Making Measurements
Measuring Amplifiers
Measuring Gain and Reverse Isolation Simultaneously
(ES Analyzers Only)
Since an amplifier will have high gain in the forward direction and high isolation in the 
reverse direction, the gain (S
21
) will be much greater than the reverse isolation (S
12
). 
Therefore, the power you apply to the input of the amplifier for the forward measurement 
(S
21
) should be considerably lower than the power you apply to the output for the reverse 
measurement (S
12
). By applying low power in the forward direction, you’ll prevent the 
amplifier from being saturated. A higher power in the reverse direction keeps noise from 
being a factor in the measurement and accounts for any losses caused by attenuators or 
couplers on the amplifier’s output needed to lower the output power into the analyzer. The 
following steps demonstrate the features that best accomplish these measurements.
1. Press 
 .
Coupling the channels allows you to have the same frequency range and calibration 
applied to channel 1 and channel 2.
2. Press 
 .
Uncoupling the port power allows you to apply different power levels at each port. In 
, the port 1 power is set to 
25 dBm for the gain measurement (S
21
) and the 
port 2 power is set to 0 dBm for the reverse isolation measurement (S
12
).
3. Press 
 
 
 
 and set the power level for 
port 1.
4. Press 
 
 
 
 and set the power level for 
port 2.
5. Perform an error-correction and connect the amplifier to the network analyzer. Refer to 
the 
 for error-correction procedures.
6. Press 
  
.
You can view both measurements simultaneously by using the dual channel display 
mode. Refer to 
. If the port power levels are in different power ranges, one of 
the displayed measurements will not be continually updated and the annotation 
tsH
 
will appear on the left side of the display. Refer to 
 for information on how to override this state.
NOTE
To obtain best accuracy, you should set the power levels prior to performing 
the calibration. However, the analyzer compensates for nominal power 
changes you make during a measurement, so that the error correction still 
remains quite valid. In these cases, the 
Cor
 annunciator will change to 
C
.
Sweep Setup
 COUPLED CH ON 
Power
 PORT POWER [UNCOUPLED] 
Chan 1
Meas
 Trans:FWD S21 (B/R) 
Power
Chan 2
Meas
 Trans: REV S12 (A/R) 
Power
Display
 DUAL | QUAD SETUP   DUAL CHAN ON