Agilent Technologies 20ET ユーザーズマニュアル

ページ / 478
1-68
 
Making Measurements
Using the Swept List Mode to Test a Device
2. Set the following measurement parameters:
 
 or on ET models: 
 
 
 
 
 
Observe the Characteristics of the Filter 
Figure 1-55 Characteristics of a Filter 
• Generally, the passband of a filter exhibits low loss. A relatively low incident power may 
be needed to avoid overdriving the next stage of the DUT (if that stage contains an 
amplifier) or the network analyzer receiver.
• Conversely, the stopband of a filter generally exhibits high isolation. To measure this 
characteristic, the dynamic range of the system will have to be maximized. This can be 
done by increasing the incident power and narrowing the IF bandwidth.
Choose the Measurement Parameters 
1. Decide the frequency ranges of the segments that will cover the stopbands and 
passband of the filter. For this example, the following ranges will be used:
• Lower stopband: 650 to 880 MHz
• Passband: 880 to 920 MHz
• Upper stopband: 920 to 1150 MHz
2. To set up the swept list measurement, press:
 
 
Meas
 Trans: FWD S21 (B/R) 
 TRANSMISSN 
Center
900
M/
µ
Span
500
M/
µ
Sweep Setup
 SWEEP TYPE MENU   EDIT LIST