Agilent Technologies 20ET ユーザーズマニュアル

ページ / 478
1-70
 
Making Measurements
Using the Swept List Mode to Test a Device
8. To maximize the dynamic range in the stopband (increasing the incident power and 
narrowing the IF bandwidth), press:
 
 
 
 
 
9. Press 
 .
Calibrate and Measure 
1. Remove the DUT and perform a full two-port calibration. Refer to 
2. With the thru connected, set the scale to autoscale to observe the benefits of using 
swept list mode.
• The segments used to measure the stopbands have less noise, thus maximizing 
dynamic range within the stopband frequencies.
• The segment used to measure the passband has been set up for faster sweep speed 
with more measurement points.
Figure 1-56 Calibrated Swept List Thru Measurement 
3. Reconnect the filter and adjust the scale to compare results with the first filter 
measurement that used a linear sweep.
In 
, notice that the noise level has decreased over 10 dB, confirming that the 
noise reduction techniques in the stopbands were successful. Also, notice that the 
stopband noise in the third segment is slightly lower than in the first segment. This is 
due to the narrower IF bandwidth of the third segment (300 Hz).
 MORE 
 SEGMENT POWER 
10
x1
 SEGMENT IF BW 
300
x1
 RETURN   DONE 
 DONE  LIST FREQ [SWEPT]