NI 5412 Specifications
Maximum Output Voltage
Maximum
Output
Voltage
Output
Voltage
Path
Load
Maximum Output Voltage (V
pk-pk
)
The Maximum
Output Voltage
of the NI 5412 is
determined by
the Amplitude
Range and the
Offset Range.
Output Voltage
of the NI 5412 is
determined by
the Amplitude
Range and the
Offset Range.
Low-
Gain
Amplifier
50
Ω
±1.000
1 k
Ω
±1.905
Open
±2.000
High-
Gain
Amplifier
50
Ω
±6.000
1 k
Ω
±11.43
Open
±12.00
Accuracy
DC Accuracy
±0.2% of Amplitude ± 0.05% of Offset ± 500 µV
(within ±10 °C of self-calibration temperature)
(within ±10 °C of self-calibration temperature)
±0.4% of Amplitude ± 0.05% of Offset ± 1 mV
(0 °C to 55 °C)
(0 °C to 55 °C)
Calibrated for
high impedance
load.
high impedance
load.
AC Amplitude
Accuracy
Accuracy
±1.0% of desired Amplitude ± 1 mV
50 kHz sine
wave.
wave.
Output Characteristics
Output
Impedance
Impedance
50
Ω or 75 Ω nominal, software-selectable.
—
Output
Coupling
Coupling
DC
—
Output Enable
Software-selectable. When the Output Path is disabled, the
CH 0 Output is terminated to ground with a 1 W resistor
equal to the selected output impedance.
CH 0 Output is terminated to ground with a 1 W resistor
equal to the selected output impedance.
—
Maximum
Output
Overload
Output
Overload
The CH 0 output can be connected to a 50
Ω, ±12 V source
without sustaining any damage. No damage occurs if the
CH 0 output is shorted to ground indefinitely.
CH 0 output is shorted to ground indefinitely.
—
Waveform
Summing
Summing
The CH 0 output supports waveform summing among
similar paths—specifically, the outputs of multiple NI 5412
signal generators can be connected directly together.
similar paths—specifically, the outputs of multiple NI 5412
signal generators can be connected directly together.
—
Table 1. (Continued)
Specification
Value
Comments