NI 5412 Specifications
Maximum Output Voltage
Maximum 
Output 
Voltage
Path
Load
Maximum Output Voltage (V
pk-pk
)
The Maximum 
Output Voltage 
of the NI 5412 is 
determined by 
the Amplitude 
Range and the 
Offset Range.
Low-
Gain 
Amplifier
50 
Ω
±1.000
1 k
Ω
±1.905
Open
±2.000
High-
Gain 
Amplifier
50 
Ω
±6.000
1 k
Ω
±11.43
Open
±12.00
Accuracy
DC Accuracy
±0.2% of Amplitude ± 0.05% of Offset ± 500 µV 
(within ±10 °C of self-calibration temperature)
±0.4% of Amplitude ± 0.05% of Offset ± 1 mV 
(0 °C to 55 °C)
Calibrated for 
high impedance 
load.
AC Amplitude 
Accuracy
±1.0% of desired Amplitude ± 1 mV
50 kHz sine 
wave.
Output Characteristics
Output 
Impedance
50 
Ω or 75 Ω nominal, software-selectable.
Output 
Coupling
DC
Output Enable
Software-selectable. When the Output Path is disabled, the 
CH 0 Output is terminated to ground with a 1 W resistor 
equal to the selected output impedance.
Maximum 
Output 
Overload
The CH 0 output can be connected to a 50 
Ω, ±12 V source 
without sustaining any damage. No damage occurs if the 
CH 0 output is shorted to ground indefinitely.
Waveform 
Summing
The CH 0 output supports waveform summing among 
similar paths—specifically, the outputs of multiple NI 5412 
signal generators can be connected directly together.
Table 1.   (Continued)
Specification
Value
Comments