C O N T R O L   P A N E L S  
 
Display Software Version 
To display the software version, enter the configuration mode (See instructions listed in the 
Configuration Mode – cnf section listed above). When cnf is displayed, press the DOWN 
button. The software version number will be displayed and should look something like 
d.
17
Press the DOWN arrow key again to return back to the cnf display. 
Exit the supervisor mode when done. 
NOTE:
 See Supervisor Mode section on how to exit. 
Alarms 
Alarm 1 is associated with the heater’s sheath temperature, and is triggered when the 
temperature rises above the value set in al.1 (see chart above). 
Alarm 2 is associated with the wash tanks water temperature, and is triggered when the 
temperature rises above the value set in al.2 (see chart above). 
When one of these alarms is activated, the circulation pump will turn off and the alarm 
indicator light will be illuminated. The controller will enter the fil state. 
 
Display Sheath Temperature 
During normal operation, pressing the UP button will display the heater’s sheath 
temperature. The sheath temperature will be displayed for about six seconds and then switch 
back to the original display that was shown before the UP button was pressed. 
 
Reset Change Water 
The Change Water flag is cleared whenever the heaters sheath temperature rises above its set 
point. In the case of a heater failure this flag cannot be reset automatically. To manually reset 
this flag you will need to hold down the UP button and START/STOP button 
simultaneously for approximately 15 seconds. The controller will then enter the fil state. 
 
Display Lamp Test 
A lamp test can be started by pressing the DOWN and TEMP buttons simultaneously. 
During the lamp test, the controller functions normally, but all display segments and status 
lights flash. Pressing the DOWN and TEMP buttons again will return back to the normal 
operation display. 
 
Temperature Sensor Fault 
If the water or heater sheath temperature sensors should fail, the controller will indicate the 
fault by taking the controller to the fil state. The alarm light will be on, the heater will be 
disabled, and the circulation pump will be off. 
After the sensor(s) have been replaced, normal operation is resumed by exiting the fil state. 