6.5.3  Gain Calibration 
1  Put a collimator on center of the detector, and open the collimator completely. 
2  Adjust SID as 130㎝  ~ 150㎝  to make X-ray exposure range include the detector. 
3  Remove any objects or foreign materials between the tube and the detector. 
 
      
 
 
4  Choose Gain Calibration -Normal
5  Set the tube voltage of X-ray generator as 70㎸  ~ 80㎸
6  While acquiring images, adjust the X-ray dose condition (  or  ) until Current Value of the Gain 
Calibration area is formed between 8000±200
7  Set the exposure number of Stage as 10 when the adjustment of exposure condition is completed. 
8  Click Get button, and keep making an X-ray exposure at 15 sec intervals. 
9  Save the Gain Calibration data as a file. 
 
 
 
 
 
Make sure to progress Gain Calibration again if the cases below are applicable. 
 
When the settings of Exposure Section or Gain Type is changed from the VIVIX Setup 
program. 
 
When the exposure devices like X-ray generator are repaired. 
 
When the exposure environment is changed. 
 
When the service engineer decides that it is need to progress Gain Calibration.