VIVIX-S 1012N User Manual 
6.5.4  Auto Defect Correction 
 
 
While operating the detector, defect pixels may appear on the image. In this case, use the 
defect auto correction to calibrate the defect pixels. 
 
 
 
Be sure to check the followings before acquiring the FLAT image. 
 
Preheat the detector for 30 minutes or more. 
 
The recommend SID is 150 ㎝  (distance between X-ray tube and detector) 
 
Open the collimator of X-ray tube completely. 
 
Align the center of the detector with the center of collimator. 
 
Keep everything away from the detector surface. 
 
1  Make an exposure of FLAT images from the Image dialog box. 
2  Click Save Image button to save images to the user-defined folder. (saved as a raw file.)