Rechargeable Stimulator System
33
IPG Battery Status
When the patient Remote Control communicates with the Stimulator, 
the battery status is sent to the Remote Control, which will display a 
message when the Stimulator battery is empty. When the Remote 
Control indicates a low battery (message: Recharge Stimulator Soon), 
the Stimulator should be recharged as soon as possible. Failure to 
recharge will lead to loss of stimulation in less than 24 hours, and the 
Stimulator will need to be charged for approximately two hours 
before reactivation.
If the patient reports that stimulation has stopped, instruct the patient 
to check the Stimulator battery status with the Remote Control.