KROHNE FMCW80G74TA ユーザーズマニュアル

ページ / 126
 OPERATION 
6
83
OPTIWAVE 7500 C
www.krohne.com
10/2016 - 4004375401 - MA OPTIWAVE 7500 R01 en
Procedure
Screen
Steps
Description
2 ×  [>>>>], 2 × [
] and [>>>>].
Default screen.
Enter Program Mode and go to menu item 
A3 Login.
2 ×  [>>>>], 5 × [
, [>>>>], 8 × [
] and [^
^
^
^]
Enter the password used at this time for the 
"Expert" security level. If it is the default 
password, refer to 
Protection of the device 
settings (security levels)
 on page 47.
[
], [>
>>>], [
] and 2 × [>>>>].
[
] or [
] for the selection of the 
empty spectrum type (Full 
(Average), Full (Max), Partial 
(Average), Partial (Max)).
[^
^
^
^] to confirm.
Push these buttons to start the Record 
Spectrum (Empty Spectrum) procedure.
Empty spectrum type. Make a selection 
from the list of parameters. If you can empty 
the tank, set this menu item to "Full, 
Average" or "Full, Max". If you cannot empty 
the tank, set this menu item to "Partial, 
Average" or "Partial, Max".
.
[>>>>] to change the position of the 
cursor.
[
] to decrease the value (or move 
the decimal point one digit to the 
left) or [
] to increase the value (or 
move the decimal point one digit to 
the right).
[^
^
^
^] to confirm.
Partial distance. If you set Empty Spectrum 
Empty Spectrum 
Empty Spectrum 
Empty Spectrum 
Type
Type
Type
Type to "Partial, Average" or "Partial, Max", 
you will have one more step to do in this 
procedure. You must give the distance 
(partial distance) of the product surface 
from the flange facing or thread stop of the 
process connection.
[^
^
^
^] to confirm.
Make sure that the tank is empty or it 
contains not more than the minimum 
quantity. .
[
] or [
] to change the parameter 
("No" or "Yes".
[^
^
^
^] to confirm.
Start recording? Set this menu item to "Yes" 
to continue to the subsequent step. Set this 
menu item to "No" to go back to the menu.