OPTIWAVE 7500 C
www.krohne.com
10/2016 - 4004375401 - MA OPTIWAVE 7500 R01 en
Procedure
Screen
Steps
Description
•
2 × [>>>>], 2 × [
] and [>>>>].
Default screen.
Enter Program Mode and go to menu item
A3 Login.
•
2 × [>>>>], 5 × [
, [>>>>], 8 × [
] and [^
^
^
^]
Enter the password used at this time for the
"Expert" security level. If it is the default
password, refer to
Protection of the device
settings (security levels)
on page 47.
•
[
], [>
>>>], [
] and 2 × [>>>>].
•
[
] or [
] for the selection of the
empty spectrum type (Full
(Average), Full (Max), Partial
(Average), Partial (Max)).
(Average), Full (Max), Partial
(Average), Partial (Max)).
•
[^
^
^
^] to confirm.
Push these buttons to start the Record
Spectrum (Empty Spectrum) procedure.
Empty spectrum type. Make a selection
from the list of parameters. If you can empty
the tank, set this menu item to "Full,
Average" or "Full, Max". If you cannot empty
the tank, set this menu item to "Partial,
Average" or "Partial, Max".
.
•
[>>>>] to change the position of the
cursor.
cursor.
•
[
] to decrease the value (or move
the decimal point one digit to the
left) or [
left) or [
] to increase the value (or
move the decimal point one digit to
the right).
the right).
•
[^
^
^
^] to confirm.
Partial distance. If you set Empty Spectrum
Empty Spectrum
Empty Spectrum
Empty Spectrum
Type
Type
Type
Type to "Partial, Average" or "Partial, Max",
you will have one more step to do in this
procedure. You must give the distance
(partial distance) of the product surface
from the flange facing or thread stop of the
process connection.
•
[^
^
^
^] to confirm.
Make sure that the tank is empty or it
contains not more than the minimum
quantity. .
•
[
] or [
] to change the parameter
("No" or "Yes".
•
[^
^
^
^] to confirm.
Start recording? Set this menu item to "Yes"
to continue to the subsequent step. Set this
menu item to "No" to go back to the menu.