Endress and Hauser GmbH and Co FMR6XET ユーザーズマニュアル

ページ / 88
Micropilot FMR62
Endress+Hauser
V. 4, Rev. 2, 19-03-2018
65
3A and EHEDG approval for versions with Tri-Clamp and DIN11851 process connections.
To avoid any risk of contamination, install the device according to the “Hygienic Equipment
Design Criteria (HDC)” which was published by the Subgroup "Design Principles" of the EHEDG
on April 8 2004.
Suitable fittings and seals must be used to ensure hygiene-compliant design according to the
specifications of 3-A SSI and EHEDG.
The gap-free connections can be cleaned of all residue using the usual cleaning methods within
this industry.
The 316L housing (GT18), option B in order feature 40, is not suitable for use in applications
which require a 3A approval.
NACE MR 0175 / ISO 15156
• The wetted, metal materials comply with the requirements of NACE MR 0175 / ISO 15156.
• Declaration of Conformity: see the product structure, feature 580, version JB
NACE MR 0103
• The wetted, metal materials comply with the requirements of NACE MR 0103.
• The Declaration of Conformity is based on NACE MR 0175.
The hardness and intergranular corrosion have been tested, and heat treatment (solution
annealed) has been performed. The materials used therefore meet the requirements of
NACE MR 0103.
• Declaration of Conformity: see the product structure, feature 580, version JE.
Pressure equipment with
allowable pressure
≤ 200 bar (2
 
900 psi)
Pressure instruments with a flange and threaded boss that do not have a pressurized housing do not
fall within the scope of the Pressure Equipment Directive, irrespective of the maximum allowable
pressure.
Reasons:
According to Article 2, point 5 of EU Directive 2014/68/EU, pressure accessories are defined as
"devices with an operational function and having pressure-bearing housings".
If a pressure instrument does not have a pressure-bearing housing (no identifiable pressure chamber
of its own), there is no pressure accessory present within the meaning of the Directive.
Marine approval
Device
Marine approval
LF:
ABS marine
approval
LG:
LR marine approval
LH:
BV marine approval
LJ:
DNV GL marine
approval
FMR62
1)
See order code for 590 "Additional approval"
Radio standard
EN 302729-1/2
The devices FMR62 with antenna 40 mm / 2" integrated PEEK, comply with the Level Probing Radar
(LPR) radio standard EN 302729-1/2. The devices FMR62 with antenna 40 mm / 2" integrated
PEEK, are approved for unrestricted use inside and outside closed containers in countries of the EU
and the EFTA . that have already implemented this standard.
The following countries are those that have currently implemented the directive:
Belgium, Bulgaria, Germany, Denmark, Estonia, France, Greece, UK, Ireland, Iceland, Italy,
Liechtenstein, Lithuania, Latvia, Malta, The Netherlands, Norway, Austria, Poland, Portugal,
Romania, Sweden, Switzerland, Slovakia, Spain, Czech Republic and Cyprus.
Implementation is still underway in all of the countries not listed.
Please note the following for operation of the devices outside of closed vessels:
1. Installation must be carried out by properly trained, expert staff.
2. The device antenna must be installed in a fixed location pointing vertically downwards.
6)
EHEDG pending