Appendix: Electrical Safety 
Aborting a Test 
± Deactivating the Capsule 
Caution 
If a test is aborted before the capsule has been exposed to buffer solution, the capsule may be deactivated but 
must be used within two (2) hours or discarded. 
Caution 
If a test is aborted before ingestion and the capsule has not been exposed to buffer solution, the capsule may 
be deactivated but must be used within one month or before its expiration date is reached, whichever is 
sooner.  Otherwise discard the capsule (refer to Recycling and Disposal Instructions, page 18). 
1. Place the capsule’s container onto the activation fixture.  
2. Align the diamond-shaped window on the capsule’s container with the Off mark 
on the activation fixture.  
3. Leave the capsule container in place for 5 seconds and then lift straight up and off 
the activation fixture. 
4. Confirm the capsule is deactivated: 
x Place the capsule container in the docking station’s well and observe the data 
receiver’s display. The capsule is off if this icon   (lower right) is replaced with 
an X.  The X may take 1-2 minutes to appear.