Given Imaging Limited SMARTPILL ユーザーズマニュアル

ページ / 120
Appendix: Electrical Safety 
x Select Action > Download Data. 
x Click the 
 toolbar icon. 
The Patient Information screen opens. 
 
Post Test Notes 
7. In the Procedure Notes tab, click in the post-test notes field to enter notes elicited 
from the patient which were captured when the patient returned the data receiver. 
x See Adding and Creating a Template.  
8. Click OK. (If you click Cancel, the data receiver continues to record capsule data 
and events.) 
The Test Download screen appears. 
Duration of test 
Duration of download (typical) 
2 days 
<3 minutes 
4 days 
<5 minutes 
MotiliGI protects current test data stored on the data receiver until current test data 
is completely downloaded. 
When the download is complete the MotiliGI software displays the entire test 
graph and the Test Summary screen. 
9. Turn the data receiver off and place in the docking station to be recharged.