Boston Scientific Corporation CRM312004 ユーザーズマニュアル
12
Figure 6. Model 3120 PRM right-side keypad.
Monitoring/Recording Functions
Surface ECG
Pressing any speed key will run a surface ECG if the surface ECG patient
cable is connected to the PRM and electrodes are attached to the patient
(unless a report is being printed). For detailed instructions, refer to “ECG
Display” on page 10.
cable is connected to the PRM and electrodes are attached to the patient
(unless a report is being printed). For detailed instructions, refer to “ECG
Display” on page 10.
Press STAT SHOCK when in telemetry
communication with the pulse generatorto program
predetermined high-energy parameters for
emergency shock delivery.
communication with the pulse generatorto program
predetermined high-energy parameters for
emergency shock delivery.
Press PROGRAM when in telemetry communication
with the pulse generator to transmit new parameter
values to the pulse generator.
with the pulse generator to transmit new parameter
values to the pulse generator.
Press INTERROGATE when in telemetry
communication with the pulse generator to obtain
information stored in the pulse generator’s memory.
communication with the pulse generator to obtain
information stored in the pulse generator’s memory.
Table 1 Key Functions
STAT SHOCK
PROGRAM
INTERROGATE
STAT PACE
STAT SHOCK
DIVERT THERAPY
PROGRAM
INTERROGATE