Section 8:  MAINTENANCE 
 
Annual Maintenance 
The 
EPIC 10
 should be serviced annually by a qualified, trained, and certified technician.  
Annual calibrations can be performed at a certified depot repair facility. Call Biolase Service 
at 1-800-321-6717 or your Authorized Service Representative to schedule an appointment. 
Please contact Biolase Service at 1-800-321-6717 or your Authorized Representative to 
discuss Extended Service Contracts and Annual maintenance options. 
 
Daily Maintenance 
Use peel-off clear covers supplied with the system.  Use disinfectant to wipe down the front 
panel and handpiece holder of the 
EPIC 10
 system after each procedure.  Do not use 
bleach or abrasive cleansers. 
Check and clean the protective window of the fiber optic shaft with cotton swab wet in 
alcohol. 
 
Contamination Control Procedures 
The contamination control suggested for the 
EPIC 10
 surgical handpiece and tips is the 
steam sterilization method.  However, before sterilization, the 
EPIC 10
 reusable handpiece 
should be carefully cleaned per the following procedure. 
 
NOTE:
  The fiber optic cable and Handpiece are delivered from the manufacturer as 
 
              non-sterile.
 
NOTE:
  Tips are designed to withstand a single sterilization cycle and must be
 
              disposed of after single use in a biohazard medical waste sharps container.
 
 
Cleaning Instructions for the Surgical Handpiece, the Classical (Feed-Through) Fiber 
Handpiece, and the Reusable Fiber Optic Cable 
The cleaning process is intended to remove blood, protein and other potential contaminants 
from the surfaces and crevices of reusable accessories.  This process may also reduce the 
quantity of particles, microorganisms and pyrogens present. Cleaning should be performed 
prior to sterilization and must be conducted only by qualified office personnel trained to 
perform the procedure and handle the  
EPIC 10
 Fiber optic Delivery System. 
Wear protective latex gloves when handling the contaminated delivery system.  
To clean the fiber cable, wipe the entire cable, including the shaft, with cotton gauze and 
chemical disinfectant.  Keep the window intact.  If the window is dirty, clean with a cotton 
swab wet with alcohol.