Biolase Inc. EPIC-1 ユーザーズマニュアル

ページ / 56
EPIC™10 User Manual 
pg. 41 
5400525 Rev B  
 
 
 
Title 
Message  
  
Reason   
 
Fix   
 
Error 1 
 
Thermistor Open 
Thermistor Open 
Call BIOLASE Service  
Error 2 
 
Thermistor Shorted 
Thermistor Shorted 
Error 3 
 
Shutdown Temperature 
System too hot 
Allow 5-10 mins for laser to cool 
down 
Error 4 
 
Laser Current High/ Low 
Output is out of specs 
Call BIOLASE Service  
Error 5 
 
FS shorted in Standby 
FS is partially pressed or 
is damaged 
Press/Release FS or call Biolase 
Service 
Error 6 
 
ON/OFF button Stuck 
Key stuck 
Press Front key 
Error 7 
 
Flash Corrupted 
Memory Corrupted 
Call BIOLASE Service  
Error 8 
 
No Fiber 
Fiber not inserted 
Plug in Trunk Fiber 
Error 9 
 
Lost Footswitch Communication  Wireless Interference 
Reposition console or FS to 
improve communication 
Error 10 
 
Emergency Switch 
E-Switch Pressed 
Press E-Switch Again 
Error 11 
 
Remote Interlock 
Remote interlock open 
Check Remote Interlock closed 
Error 12 
 
Battery Critically Low 
Battery Critically Low 
Plug in DC supply 
Warning 1 
 
Temp High 
System is hot 
Allow 5-10 mins for laser to cool 
down 
Warning 2 
 
Battery Low 
Battery is low 
Plug in DC supply 
Warning 3 
 
Battery Not Connected 
Battery not connected 
Plug in Battery 
Warning 4 
 
FS Battery Low 
Battery on FS low 
Replace FS battery 
Alert 1 
 
Wireless Not Paired 
No wireless connect 
Re-establish pairing (see Sec 4) 
Alert 2 
 
System must be in READY 
mode to lase 
System is not in READY 
mode 
Press the Control Button in any 
procedure screen 
Figure 11.1