APPENDIX E – ELECTROMAGNETIC COMPATIBILITY
GUIDANCE AND MANUFACTURER’S DECLARATION – ELECTROMAGNETIC IMMUNITY
The model Epic 10 is intended for use in the electromagnetic environment specified below. The customer
The model Epic 10 is intended for use in the electromagnetic environment specified below. The customer
or the user of the model Epic 10 should assure that it is used in such an environment.
Immunity test
IEC 60601
test level
Continuous
level
Electromagnetic environment -
guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
± 6 kV contact
± 8kV air
± 6 kV contact
± 8kV air
Floors should be wood, concrete or
ceramic tile. If floors are covered with
synthetic material, relative humidity
should be at least 50%.
Electrical fast
transient/burst
IEC61000-4-4
± 2 kV for power
supply lines
± 1 kV for
input/output lines
± 2 kV for power
supply lines
N/A
Main power quality should be that of a
typical commercial or hospital
environment.
Input/output that does not apply
because the footswitch cable length is
less than 3 meters.
Surge
IEC 61000-4-5
± 1 kV differential
mode
± 2kV common
mode
± 1 kV differential
mode
± 2kV common
mode
Mains power quality should be that of
a typical commercial or hospital
environment.
Voltage dips, short
interruptions and
voltage variations
on power supply
input lines.
IEC 61000-4-11
<5% U
r
(>95% dip in UT)
for 0.5 cycle
40% U
r
(60% dip in UT) for
5 cycles
70% U
r
(30% dip in U
r
) for
25 cycles
<5% Ur
(>95% dip in U
(>95% dip in U
r
) for
5 seconds
<5% U
r
(>95% dip in UT)
for 0.5 cycle
40% U
r
(60% dip in UT) for
5 cycles
70% U
r
(30% dip in U
r
) for
25 cycles
<5% Ur
(>95% dip in U
(>95% dip in U
r
) for
5 seconds
Mains power quality should be that of
a typical commercial or hospital
environment. If the user of the model
Epic 10 requires continued operation
during power mains interruptions, it is
recommended that the model Epic 10
be powered from an uninterrupted
power supply.
Power frequency
(50-60 Hz)
magnetic field
IEC 61000-4-8
3 A/m
3 A/m
Power frequency magnetic fields
should be at levels characteristic of a
typical location in a typical commercial
or hospital environment.
NOTE: U
r
is the A.C. mains voltage prior to applications of the test level.