GUIDANCE AND MANUFACTURER’S DECLARATION – ELECTROMAGNETIC IMMUNITY     (Continued) 
The model Epic 10 is intended for use in the electromagnetic environment specified below.  The customer 
or the user of the model Epic 10 should assure that it is used in such an environment. 
Immunity test 
IEC 60601  
test level 
Continuous 
 level 
Electromagnetic environment - 
guidance 
Conducted RF  
 
IEC 61000-4-6 
 
 
Radiated RF 
 
IEC61000-4-3 
3 Vrms 
 
150 kHz to 80 GHz 
 
 
3V/m 
 
80 MHz to 2.5 GHz 
3 V 
 
3Vm 
Portable and mobile RF 
communications equipment should be 
used no closer to any part of the 
model Epic 10, including cables, than 
the recommended separation distance 
calculated from the equation 
applicable to the frequency of the 
transmitter. 
 
Recommended separation distance 
 
d = 1.2√P 
 
d = 1.2√P  80 MHz to 800 MHz 
 
d = 2.3√P  800MHz to 2.5GHZ 
 
Wher P is the maximum output power 
rating of the transmitter in watts (W) 
according to the transmitter 
manufacturer and d 8s the 
recommended separation distance in 
meters (m). 
 
Field strengths from fixed RF 
transmitters, as determined by an 
electromagnetic site survey,
a
 should 
be less than the compliance level in 
each frequency range.
 
Interference may occur in the vicinity 
of equipment marked with the 
following symbol: 
           
 
NOTE 1 -  At 80 MHz and 800 MHz, the higher frequency range applies. 
NOTE 2 – These guidelines may not apply in all situations.  Electromagnetic propagation is affected by 
                  absorption and reflection from structures, objects, and people. 
A.  Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephone and land mobile radios, 
amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy.  To assess the 
electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered.  If the measured 
field strength in the location in which the Epic 10 is used exceeds the applicable RF compliance level above, the Epic 10 should be 
observed to verify normal operation.  If abnormal performance is observed, additional measures may be necessary, such as 
reorienting or relocating the Epic 10. 
B.  Over the frequency range 150 kHz to 80 MHz, field strengths should be less than [V1] V/m.