Biolase Inc. EPIC-1 ユーザーズマニュアル
EPIC™10 User Manual
pg. 52
5400525 Rev B
GUIDANCE AND MANUFACTURER’S DECLARATION – ELECTROMAGNETIC IMMUNITY (Continued)
The model Epic 10 is intended for use in the electromagnetic environment specified below. The customer
The model Epic 10 is intended for use in the electromagnetic environment specified below. The customer
or the user of the model Epic 10 should assure that it is used in such an environment.
Immunity test
IEC 60601
test level
Continuous
level
Electromagnetic environment -
guidance
Conducted RF
IEC 61000-4-6
Radiated RF
IEC61000-4-3
3 Vrms
150 kHz to 80 GHz
3V/m
80 MHz to 2.5 GHz
3 V
3Vm
Portable and mobile RF
communications equipment should be
used no closer to any part of the
model Epic 10, including cables, than
the recommended separation distance
calculated from the equation
applicable to the frequency of the
transmitter.
Recommended separation distance
d = 1.2√P
d = 1.2√P 80 MHz to 800 MHz
d = 2.3√P 800MHz to 2.5GHZ
Wher P is the maximum output power
rating of the transmitter in watts (W)
according to the transmitter
manufacturer and d 8s the
recommended separation distance in
meters (m).
Field strengths from fixed RF
transmitters, as determined by an
electromagnetic site survey,
a
should
be less than the compliance level in
each frequency range.
b
Interference may occur in the vicinity
of equipment marked with the
following symbol:
NOTE 1 - At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 – These guidelines may not apply in all situations. Electromagnetic propagation is affected by
NOTE 2 – These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects, and people.
A. Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephone and land mobile radios,
amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the
electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured
field strength in the location in which the Epic 10 is used exceeds the applicable RF compliance level above, the Epic 10 should be
observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as
reorienting or relocating the Epic 10.
B. Over the frequency range 150 kHz to 80 MHz, field strengths should be less than [V1] V/m.