meet the requirements of those countries. 
 
 
Provide the right cord-set. In US and Canada “hospital grade” cord-set 
has to be used, provided with instructions to indicate that grounding 
reliability can be achieved only when the equipment is connected to an 
equivalent receptacle marked hospital only or hospital grade. These 
instructions need to be marked either on the equipment or on a tag on the 
power cord. 
 
 
When in use, place the AC adapter in an open and ventilated area, such as 
on a desktop or the floor (do not use on wet surface). Do not cover the AC 
adapter with items that will reduce cooling. 
Pay attention that the adapter cable and the power cord are routed far 
from places where they can be tripped over or stepped on. 
 
 
Disconnect power cords and cables by grabbing the plug and not by 
pulling on the cable itself. As you pull out the connector, keep it evenly 
aligned to avoid bending any connector pins. Before you connect a cable, 
make sure both connectors are correctly aligned. 
 
 
If you use an extension cord with your AC adapter, ensure that the total 
ampere rating of the products plugged into the extension cord does not 
exceed the ampere rating of the extension cable. 
When using your power cord, make sure to position it around objects so it 
will not be cut or punctured. 
 
 
 
 
Anybody connecting additional equipment to medical electrical equipment 
configures a medical system and is therefore responsible that the system 
complies with the requirements for medical electrical systems. Attention is 
drawn to the fact that local laws take priority over the above mentioned 
requirements. If in doubt, consult your local representative or the 
technical service department 
 
To avoid the risk of electric shock, this equipment must only be connected 
to a supply mains with protective earth 
 
 
 
 
 
The internal electrical power source has to be used if the integrity of the 
protective earth conductor or the protective earthing system in the 
installation is in doubt. 
 
 
5.12 _________________________________________________________ 
Battery 
 
The NGP1 Computing Platform comes with a lithium ion rechargeable main battery having 
a capacity of 4000mAh at an output voltage of 11.1 V (P/N 4519 208 81321). The first 
instructions on how insert and charge the battery are given in section  3.1.1  The NGP1 
Computing Platform has three battery indicators that show the charge/discharge status. 
Refers to 3.1.5.