meet the requirements of those countries.
Provide the right cord-set. In US and Canada “hospital grade” cord-set
has to be used, provided with instructions to indicate that grounding
has to be used, provided with instructions to indicate that grounding
reliability can be achieved only when the equipment is connected to an
equivalent receptacle marked hospital only or hospital grade. These
instructions need to be marked either on the equipment or on a tag on the
power cord.
instructions need to be marked either on the equipment or on a tag on the
power cord.
When in use, place the AC adapter in an open and ventilated area, such as
on a desktop or the floor (do not use on wet surface). Do not cover the AC
adapter with items that will reduce cooling.
adapter with items that will reduce cooling.
Pay attention that the adapter cable and the power cord are routed far
from places where they can be tripped over or stepped on.
from places where they can be tripped over or stepped on.
Disconnect power cords and cables by grabbing the plug and not by
pulling on the cable itself. As you pull out the connector, keep it evenly
pulling on the cable itself. As you pull out the connector, keep it evenly
aligned to avoid bending any connector pins. Before you connect a cable,
make sure both connectors are correctly aligned.
make sure both connectors are correctly aligned.
If you use an extension cord with your AC adapter, ensure that the total
ampere rating of the products plugged into the extension cord does not
ampere rating of the products plugged into the extension cord does not
exceed the ampere rating of the extension cable.
When using your power cord, make sure to position it around objects so it
will not be cut or punctured.
When using your power cord, make sure to position it around objects so it
will not be cut or punctured.
Anybody connecting additional equipment to medical electrical equipment
configures a medical system and is therefore responsible that the system
complies with the requirements for medical electrical systems. Attention is
complies with the requirements for medical electrical systems. Attention is
drawn to the fact that local laws take priority over the above mentioned
requirements. If in doubt, consult your local representative or the
technical service department
technical service department
To avoid the risk of electric shock, this equipment must only be connected
to a supply mains with protective earth
The internal electrical power source has to be used if the integrity of the
protective earth conductor or the protective earthing system in the
installation is in doubt.
installation is in doubt.
5.12 _________________________________________________________
Battery
The NGP1 Computing Platform comes with a lithium ion rechargeable main battery having