CORITICAL ACTIVITY INDEX 
|
 36
 
 
 
Chapter 3 Operation Method of CAI 
Conductive component of electrode or sensor and connector should not be in contact with any 
other conductive parts including ground. 
When using high frequency surgical device, CAIs sensor electrode should not be located 
between operation site and electrical surgical device return electrode. 
When using brain stimulation device (ex: brain current evoked potential motor), CAIs should be 
located as distant as possible to reduce risk of burning and make for the sensors to be 
attached according to attachment locations illustrated on the sensor package.  
CAIs and CAIx connection cable should be located in a stable site to avoid discomfort of 
patients.  
 
3.1  Connecting CAIs 
 
 
Figure 3.1 Connecting CAIs 
 
Step 1:
 
Place the front side of CAIs sensor (side with sensor drawing) upward.   
Step 2: 
Insert the CAIs sensor into CAIx connector until “click” is heard.  
Step 3: 
Check the instruction indicated on the package of CAIs and attach CAIs to connection and 
body area (forehead) firmly as not to be detached.