Gait Up PHYS5POMOC ユーザーズマニュアル
12 June 2017 - Physilog5usermanual.pdf-v1.2.3 - Check www.gaitup.com for the latest version. Copyright 2017 Gait Up SA. CH-1015 SWITZERLAND
2
1. Physilog®5 elements
2. Charging
The Physilog®5 is charged via the microUSB port. Plug the microUSB connector of the cable
delivered with the Physilog®5 to the Physilog®. Plug the USB connector to a Computer or
charger (see voltage specifications below). Physilog®5 must be turned off for charging. Do
not charge Physilog® when room temperature is outside of range 0°C to 45°C.
delivered with the Physilog®5 to the Physilog®. Plug the USB connector to a Computer or
charger (see voltage specifications below). Physilog®5 must be turned off for charging. Do
not charge Physilog® when room temperature is outside of range 0°C to 45°C.
Warning: Be careful to plug the microUSB cable in the correct orientation into
the Physilog®, forcing the connector may permanently damage the Physilog®
and compromise its waterproofing.
Low battery level is indicated by orange (red and green LED on) blinking during
measurement or if the Physilog® blinks rapidly 3 times red when starting and doesn't start
the recording (see LED indications below). A battery level estimation is available in the
mobile application (free to download from www.gaitup.com/support). When the Physilog®
is discharged, plug it for charging for at least 30 minutes before use.
measurement or if the Physilog® blinks rapidly 3 times red when starting and doesn't start
the recording (see LED indications below). A battery level estimation is available in the
mobile application (free to download from www.gaitup.com/support). When the Physilog®
is discharged, plug it for charging for at least 30 minutes before use.
It is recommended to avoid complete discharging of the Physilog®5, therefore check that the
Physilog® sensors are off when you store them. After a long period without using the
Physilog®, plug it for charging for at least 30 minutes before use. It may happen that the LED
doesn't blink orange at the beginning of charging after the battery was very low, to see the
charging ongoing shortly push the main button or disconnect and reconnect after some
charging time.
Supply Voltage specifications
The Physilog® 5 should be charged using USB at 4.2V, 100mA. For charging, the ambient
temperature must lie between 0°C and 45°C.
temperature must lie between 0°C and 45°C.