TECNICAS CIENTIFICAS PARA LABORATORIO S.A. T620 ユーザーズマニュアル
ISEsweat
- User’s Manual 20
4.1 Settings
SETUP (SET).
Once the equipment is ON, and after the software has been
loaded (you have to wait 10 seconds for this), the Start screen
will appear. On the lower part of the screen, you will see <SET>
This is the access portal to the Settings Menu (Setup). Using
this menu, you can choose between different Work Modes.
ISEweat
®
is designed to run the Sweat Test and to establish the chloride concentration in a
sweat sample.
To do so, ISEsweat
To do so, ISEsweat
®
has 2 work modes:
Normal Mode, designed to run the analysis or sweat test.
Quality Control Mode ( ISEtrol N or ISEtrol A ), designed to check the equipment. It
Quality Control Mode ( ISEtrol N or ISEtrol A ), designed to check the equipment. It
is recommended to use TECIL controls or any other approved QC material for this.
Work Mode Selection
Press <SET> at the Start Screen.
A menu will appear that allows you to choose between
different programs or Work Modes:
different programs or Work Modes:
QC (ISEtrol N or ISEtrol A) or Normal (Sweat Test)
Use the scroll arrows to select a program from the control
panel.
panel.
Press <ENT> when you have selected the Work Mode
you want to use.
You will see the Remote Linking menu screen, both in Quality Control and in Sweat Test
modes
modes
The factory setting of the remote number is zero. Skip
through this by pressing <ENT>