Guidance and Manufacturer’s Declaration
— Electromagnetic Immunity
IMMUNITY Test
IEC 60601
Test level
Compliance level
Electro magnetic
environment - guidance
Electrostaticdischarge
(ESD)
IEC 61000-4-2
IEC 61000-4-2
± 6 kV contact
± 8 kV air
± 8 kV air
± 6 kV contact
± 8 kV air
± 8 kV air
Floors should be wood,
concreteorceramictile.
Iffloorsarecovered
withsyntheticmaterial,
therelativehumidity
should be at least 30 %.
Electrical fast transient/
burst
IEC 61000-4-4
IEC 61000-4-4
± 2 kV for power supply
lines
± 1 kV for input/output
± 1 kV for input/output
lines
± 2 kV for power
supply lines
± 1 kV for input/
output Lines
Mains power quality
should be that of a
typical commercial or
hospital environment.
Surge
IEC 61000-4-5
IEC 61000-4-5
± 1 kV line(s) to line(s)
± 2 kV line(s) to earth
±1kVdifferential
Mode
± 2 kV common mode
Mains power quality
should be that of a
typical commercial or
hospital environment.
Voltage dips, short
interruptionsandvoltage
variationsonpower
supply input lines
IEC 61000-4-11
IEC 61000-4-11
<5 % UT
(>95 % dip in UT ) for
0.5 cycle
40 % UT
(60 % dip in UT) for 5
cycles
70 % UT
(30 % dip in UT) for 25
cycles
<5 % UT
(>95 % dip in UT)
for 5s
<5 % UT
(>95 % dip in UT) for
0.5 cycle
40 % UT
(60 % dip in UT) for 5
cycles
70 % UT
(30 % dip in UT) for 25
cycles
<5 % UT
(>95 % dip in UT)
for 5s
Mains power quality
should be that of a
typical commercial or
hospital environment. If
the user of RI Witness™
requirescontinued
operationduringpower
mainsinterruptions,it
is recommended that RI
Witness™ be powered
fromanuninterruptible
power supply or a
battery.
Power frequency
(50/60 Hz)
magneticfield
IEC 61000-4-8
(50/60 Hz)
magneticfield
IEC 61000-4-8
3A/m
3A/m
Power frequency
magneticfields
should be at levels
characteristicofa
typicallocationina
typical commercial or
hospital environment.
Note: UTisthea.c.mainsvoltagepriortoapplicationofthetestlevel.