Section 3
Safety Warnings
7
Research Instruments Ltd
3
Guidance and Manufacturer’s Declaration 
— Electromagnetic Immunity
IMMUNITY Test
IEC 60601 
Test level
Compliance 
level
Electro magnetic environment - 
guidance
Conducted RF IEC 
61000-4-6
Radiated RF IEC 
61000-4-3
3 Vrms
150 kHz to 80 MHz
3 V/m
80 MHz to 2.5 GHz
3 Vrms
3 V/m
PortableandmobileRFcommunications
equipment should be used no closer to 
any part of RI Witness™, including cables, 
than the recommended separation
distance calculated from the equation
applicable to the frequency of the 
transmitter.
Recommendedseparationdistance
d = [3.5/V 
1
]√p
d = [3.5/V
1
]√p
80MHz to 800MHz
d = [3.5/V
1
√p
800MHz to 2.5GHz
where  p is the maximum output
power rating of the transmitter in
watts (W) according to the transmitter
manufacturer and d is the recommended 
separationdistanceinmetres(m).Field
strengths from fixed RF transmitters,
as determined by an electromagnetic
site survey, 
should be less than the 
compliance level in each frequency 
range.
b
 Interference may occur in the 
vicinity of equipment marked with the 
following symbol: 
Note 1: At 80 MHz and 800 MHz, the higher frequency range applies.
Note 2:Theseguidelinesmaynotapplyinallsituations.Electromagneticpropagationisaffectedby
absorptionandreflectionfromstructures,objectsandpeople.
Note 3:Theseguidelinesmaynotapplyinallsituations.Electromagneticpropagationisaffectedby
absorptionandreflectionfromstructures,objectsandpeople.
a
Fieldstrengthsfromfixedtransmitters,suchasbasestationsforradio(cellular/cordless)telephones
and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted 
theoreticallywithaccuracy.ToassesstheelectromagneticenvironmentduetofixedRFtransmitters,
anelectromagneticsitesurveyshouldbeconsidered.Ifthemeasuredfieldstrengthinthelocationin
whichRIWitness™isusedexceedstheapplicableRFcompliancelevelabove,RIWitness™shouldbe
observedtoverifynormaloperation.Ifabnormalperformanceisobserved,additionalmeasuresmay
benecessary,suchasre-orientingorrelocatingRIWitness™.
b
Overthefrequencyrange150kHzto80MHz,fieldstrengthsshouldbelessthan[V]V/m.