Neutral electrode operation
The CQM indicator for non-split type neutral electrodes shines green if a non-split type
neutral electrode is detected.
7.3 Meaning of CQM indicator condition
The CQM indicator can take on the following conditions:
Indicator for split type neutral electrodes shines green
- CQM detects a split type neutral electrode.
- Sufficient contact between the neutral electrode and the patient’s skin.
- Activation is possible.
• Continue with the procedure.
Indicator for split type neutral electrodes shines red
During activation:
During standby:
1) Insufficient contact between the
neutral electrode and the patient’s skin.
2) The connection between the neutral
electrode and the electrosurgical
generator is interrupted.
- Activation stops automatically.
- An alarm tone sounds and the error
message E0202 “Insufficent neutral
electrode contact” is displayed.
• Follow the instructions given in the
error message.
1) No neutral electrode connected.
2) Insufficient contact between the
neutral electrode and the patient’s skin.
- Activation is disabled.
Indicator for non-split type neutral electrodes shines green
If a split type neutral electrode is
connected:
If a non-split type neutral electrode is
connected:
1) CQM cannot detect the split type
neutral electrode.
2) The split type neutral electrode has a
short circuit.
- The message box “Neutral electrode”
(Figure 7.2) is displayed.
• Tap [OK] to confirm the message.
• Replace the split type neutral
electrode.
1) CQM detects a non-split type neutral
electrode.
- Activation is possible.
• Continue with the procedure.
7.4 Correct usage of neutral electrodes
For the safe use of neutral electrodes refer to the instructions for use of the correspondent
neutral electrode and take account of the following:
Compatible neutral electrodes
The section “Compatible neutral electrodes” on page 89 lists different neutral electrodes
that are compatible for the use with the electrosurgical generator.