Neutral electrode operation
The CQM indicator for non-split type neutral electrodes shines green if a non-split type 
neutral electrode is detected.
7.3 Meaning of CQM indicator condition
The CQM indicator can take on the following conditions:
Indicator for split type neutral electrodes shines green
 - CQM detects a split type neutral electrode.
 - Sufficient contact between the neutral electrode and the patient’s skin.
 - Activation is possible.
 
• Continue with the procedure.
Indicator for split type neutral electrodes shines red
During activation:
During standby:
1) Insufficient contact between the 
neutral electrode and the patient’s skin.
2) The connection between the neutral 
electrode and the electrosurgical 
generator is interrupted.
 - Activation stops automatically.
 - An alarm tone sounds and the error 
message E0202 “Insufficent neutral 
electrode contact” is displayed.
 
• Follow the instructions given in the 
error message.
1) No neutral electrode connected.
2) Insufficient contact between the 
neutral electrode and the patient’s skin.
 - Activation is disabled.
 Indicator for non-split type neutral electrodes shines green
 If a split type neutral electrode is 
connected:
If a non-split type neutral electrode is 
connected:
1) CQM cannot detect the split type 
neutral electrode.
2) The split type neutral electrode has a 
short circuit.
 - The message box “Neutral electrode” 
(Figure  7.2) is displayed.
 
• Tap [OK] to confirm the message.
 
• Replace the split type neutral 
electrode.
1) CQM detects a non-split type neutral 
electrode.
 - Activation is possible.
 
• Continue with the procedure.
7.4 Correct usage of neutral electrodes
For the safe use of neutral electrodes refer to the instructions for use of the correspondent 
neutral electrode and take account of the following:
Compatible neutral electrodes
The section “Compatible neutral electrodes” on page 89 lists different neutral electrodes 
that are compatible for the use with the electrosurgical generator.