HydrINS 2 / HydrINS 2 Mini flow meter  
Installation Manual  
 
Page 34 / 83 
8.3  Mode change of the HydrINS 2 and HydrINS 2 Mini probes 
 
 
Potential damage to the device. The internal electronic components of the device can 
be damaged by static electricity, which might affect its performance characteristics and 
operation. 
 
The change of the probe power supply mode must be performed with care. If not, there is a risk of 
causing severe electrical damage to the flow meter. 
8.3.1  Equipment 
The following equipment is required to remove the HydrINS 2/HydrINS 2 Mini probe head:  
 
Flat head screwdriver 
 
3-mm Allen key 
 
Brand new Silicagel bag 
8.3.2  Removal 
1 - Remove the six screws from the head of the flow meter: 
 
Figure 23: Remove the screws from the head of the HydrINS 2 flow meter 
2 – Remove the cover:  
There are two possible cases: 
 
 
It  is  possible  to  switch  the  probe  from  one  mode  to  the  other  thanks  to  the  light  that  is 
present  (see  the  indication  below).  In  this  case,  after  switching,  immediately  begin 
reinstalling the reinforced ABS cover as explained below.