Valeport Ltd 0804IDU ユーザーズマニュアル

ページ / 83
 
HydrINS 2 / HydrINS 2 Mini flow meter  
Installation Manual  
 
Page 39 / 83 
8.5  Electrical power supply of Display A 
 
Risk of electrocution. Always disconnect the power supply before making any electrical 
connections. 
 
Potential damage to the device. The internal electronic components of the device can 
be damaged by static electricity, which might affect its performance characteristics and 
operation. 
 
Display A can be powered according to one of the following five modes: 
 
Internal 3.6 VDC lithium battery pack 
 
External 3.6 VDC lithium battery pack 
 
Internal 9 to 28 VDC alkaline battery pack 
 
An external 9 to 28 VDC power supply connected to the labelled 9-28 VDC terminals 
 
A 9 to 28 VDC external power supply connected to the military connector 
These different power supply modes can be implemented using the two elements illustrated below: 
 
Switch SW2 with two positions: Internal or External 
 
Connector J15