8.5.5  External power supply by military connector 
For connection to the military connector, use a CNT104/CNT105 cable, and then connect the banana 
plugs to an external power supply. 
1  –  Remove  all  connections  from  Connector  J15,  and  then  connect  the  additional  board,  Cat.  No., 
4005511  to  the  same  connector.  In  this case,  the  position  of  Switch  SW2  used  earlier  is  no  longer 
relevant. Switches SW1 and SW2 of the additional board will be used. 
2 - Create the following configuration:  
 
Switch SW1 of the additional board: MilSpec 
 
Switch SW2 of the additional board: External 
 
8.6  Electrical power supply of Display C 
 
Risk of electrocution. Always disconnect the power supply before making any electrical 
connections. 
 
Potential damage to the device. The internal electronic components of the device can 
be damaged by static electricity, which might affect its performance characteristics and 
operation. 
 
 
Display C may be powered in two different ways:  
 
Terminal:  By  external  power  supply  connected  to  the  20-36  VDC  (20-28  VDC  for  serial 
numbers below 41574) terminals, 
 
MilSpec: By power supply via the military connector and a CNT 104 or CNT 105 cable. 
Select the appropriate switch position indicated on the image below: