Greisinger T-Logg 120W / 4-20 4 - 20 mA Temperature Data Logger 600678 データシート

製品コード
600678
ページ / 8
E39.0.1X.6C-01                                                 Operating Instruction T-Logg 120                                                                                          page  2 of 8
1 General
The logger T-Logg 120 ... is designed as a cost-efficient solution for monitoring standard signals. It enables an in-
dividual programming of the recording time. The last 16,000 measuring values can be stored in the memory.
In addition, the LCD-display indicates both the temperature measured at the moment and the operating status of
the logger.
2 Required 
accessory
The USB interface of your PC is used to program, start and read out the T-Logg.
For this following accessory is required:
ƒ
 Interface 
converter 
USB 100 or USB 100 SL for direct connection to the USB port of the PC,
cable length approx. 1m.
ƒ
  MINISOFT software (version 7.18 or later / free of charge) to start the logger and read out the logger data.
Notice:
It’s also possible to use the T-Logg with the comprehensive software GSOFT 40K
 
(version 7.18 or later).
3 Safety 
instructions
This device has been designed and tested in accordance with the safety regulations for electronic devices.
However, its trouble-free operation and reliability cannot be guaranteed unless the standard safety measures and
special safety advises given in this manual will be adhered to when using the device.
1. Trouble-free operation and reliability of the device can only be guaranteed if the device is not subjected to any
other climatic conditions than those stated under ‘Specification’
To protect the battery the max. permissible storage and transport temperature of the device is +85°C.
2. General instructions and safety regulations for electric, light and heavy current plants, including domestic safety
regulations (e.g. VDE), have to be observed.
3. If device is to be connected to other devices (e.g. via PC) the circuitry has to be designed most carefully. Internal
connection in third party devices (e.g. connection GND and earth) may result in not-permissible voltages im-
pairing or destroying the device or another device connected.
4. If there is a risk whatsoever involved in running it, the device has to be switched off immediately and to be
marked accordingly to avoid re-starting.
Operator safety may be a risk if:
- there is visible damage to the device.
- the device is not working as specified.
- the device has been stored under unsuitable conditions for a longer time.
In case of doubt, please return device to manufacturer for repair or maintenance.
5. Warning:
Do not use this product as safety or emergency stop devices, or in any other application where failure of the
product could result in personal injury or material damage.
Failure to comply with these instructions could result in death or serious injury and material damage!
4 Connection
4.1 Interface 
connection
Assignment of the input jack
(front view to the pins):
interface [–]
not connected
interface [+]