Chauvin Arnoux P01157020 Simple Logger P01157020 ユーザーズマニュアル

製品コード
P01157020
ページ / 172
SIMPLE LOGGER II 
 
 
 
English 
 
 
Page 52 
 
MODEL 
L702 
Electrical 
Channels 
Two 
Input 
Temperature Sensor 
Humidity Sensor 
Range 
14° to 122°F (-10° to 50°C) 
5 to 85% RH 
Accuracy 
±(1% of Reading + 1°F/C) 
±(3% of Reading + 2 cts) 
Resolution 
0.1°F/C 
0.1% RH 
Sample Rate 
Maximum of 1 every 5 seconds 
Storage Rate 
Programmable from once every 5 s to 1 per day 
Storage Modes 
Start/Stop, FIFO and Extended Recording Mode* (XRM™) 
Recording Length 
15 minutes to 8 weeks, programmable using DataView
®
 
Memory 
240,000 measurements (512 KB). 
Recorded data is stored in non-volatile memory and retained 
even if battery is low or removed. 
Communication 
USB 2.0 optically isolated 
Power Source** 
2 x 1.5 V AA-cell alkaline batteries 
Battery Life 
100 hours to > 45 days (dependent on storage rate/recording length) 
Mechanical 
Dimensions 
4.94 x 2.75 x 1.28" (136 x 70 x 32 mm) w/o Sensor 
Weight (with battery) 
6.4 oz (180 g) 
Case 
UL94-V0 
Vibration 
IEC 68-2-6 (1.5 mm, 10 to 55 Hz) 
Shock 
IEC 68-2-27 (30 G) 
Drop 
IEC 68-2-32 (1 m) 
Environmental 
Operating Temperature 
14° to 122°F (-10° to 50°C) 
Storage Temperature 
-4° to 140°F (-20° to 60°C) 
Relative Humidity 
Up to 85% at 95°F (35°C), Non-condensing 
Altitude 
2000 m 
Safety & Electro-magnetic compatibility 
Safety Rating 
EN61010-1; 50 V CAT III; Pollution Degree 2 
Protection Degree 
IP40 
CE Approved 
Yes 
*See the end of chapter