H55.0.21.6C-02 
 Operating Manual GPB 3300 
page 4 of 7 
 
 
_____________________________________________________ 
_____________________________________________________________________________ 
 
5  Operation 
5.1  Display elements 
 
1:  Main display 
Display of current measuring value or of 
minimal measured value (“Lo” blinking) or 
maximal value (“Hi” blinking)  
LOBAT 
Indicates low battery
 
5.2  Pushbuttons 
 
on/off: 
on/off switch,  press key long to switch device off 
mode: 
press short: changes between actual measuring 
value, the minimal measured value (“Lo” blinking) and 
maximal measured (“Hi” blinking) value.  
press >2s: reset Min and Max value 
zero: 
press >2s: 
activates / deactivates the “Zero-function“ 
 
6  Starting 
Switch the device on with button 
 . 
 
After the segment test 
 
(                ) 
the device displays some information to its current configuration: 
 
, if offset correction is active 
, if scale corrections active 
Unit of meas. value: nbAr (=mbar) | nnHg (=mmHg) 
, if sea-level correction is active 
P.OFF, if auto power-off is active 
 
Afterwards the device is ready for measuring. 
7  Min-/max- value memory 
Display MIN-value (Lo): 
Press button mode shortly 
Display changes between „Lo‟ and min-value  
Display MAX-value (Hi): 
Press mode shortly again 
Display changes between „Hi‟ and max-value 
Display actual value again: Press mode shortly again 
Actual value is displayed 
Delete MIN/MAX: 
Press button mode for 2 s 
MIN&MAX are deleted. „CLr‟ (Clear) is displayed  
  
 
shortly 
The Min-/max- display is automatically set to actual value display after 10 s. 
The Min-/max- values are deleted if the device is switched on after it had been turned off. 
8  Zero-function (relative measuring) 
With help of the Zero-
function relative measurements can be realized: press “zero“-button for 2s.  
“ NULL” is displayed shortly and the displayed value is set to 0. Press again “zero” for 2s to display 
the absolute value again.  
Note: Setting and resetting the Zero function will delete your min-/max- value memory.  
Field of application: Over/ Under pressure measurement,  please mind the measuring range! 
9  General information on absolute pressure measurement  
The device measures the absolute pressure. This must not be confused with the pressure at sea 
level display by weather stations. There the pressure loss due to height above sea level is 
deducted. The device is also able to perform this sea-level correction on the displayed air pressure. 
Therefore 
 (Sea Level correcti
on) has to be activated (= “on”) and the current height above 
sea level (ALTI = Altitude in [m]) has to be entered at the configuration.