 Operating Manual GPB 3300 
page 5 of 7 
 
 
_____________________________________________________ 
_____________________________________________________________________________ 
 
10 Configuration of the device 
Follow these instruction to configure the functions of the device:  
  Switch the device off. 
  Press mode button and keep it pressed. Turn the device on. Release the mode button after the 
segment test (             ) finished. The first parameter “P.OFF“ is displayed. 
 
  If a parameter should be changed press key up or down (                   ) the present setting of the 
parameter will be shown and can be changed via up/down keys.  
The parameter setting is confirmed by pressing 
 . 
  Jump to the next parameter with 
 . 
Parameter 
value 
information 
button  
buttons  
 
P.OFF 
Auto Power-Off (turn-off delay)  factory setting: 20 min.
 
1 … 120 
Auto Power-Off (turn-off delay) in minutes. If no key is pressed for the 
time adjusted in this parameter, the device is automatically switched off to 
safe battery power. (adjustable range 1 to 120 min) 
OFF 
automatic power-off is deactivated (continuous operating) 
UNIT 
Unit and range of display   
factory setting: nbar = mbar
 
nbAr 
300.0 … 1100.0 mbar, resolution 0.1 mbar 
nnHG 
225.0 … 825.0 mmHG, resolution 0.1 mmHG 
Sea-level correction    
factory setting
: off 
OFF 
Sea-level correction off 
ON 
Sea-level correction on 
ALTI 
Altitude input for sea-level correction    
factory setting
: 340 
-
500 … 9000 
-
500 … 9000 m selectable 
INIT 
Restore to factory settings 
NO 
Settings are kept 
YES 
Restoring device to factory settings 
  
Pressing              again stores the settings, the instruments restarts (segment test) 
Please note:   
If there is no key pressed within the menu mode within 2 minutes, the 
configuration will be cancelled, the entered settings are lost!