16 Error and System Messages 
Display 
Description 
What to do? 
 
or
 
Err.9
 
No sensor connected 
Switch off device and connect sensor 
Connected sensor or device defective 
If 2nd sensor available, check if device is ok. Return 
defective device/sensor to manufacturer for repair 
Value extremely out of measuring range 
Check: pressure not within sensor range? 
No display or 
confused 
characters, 
 
device does not 
react on keypress
 
Battery empty 
Replace battery 
Mains operation: wrong voltage or polarity 
Check  power supply, replace it when necessary 
System error 
Disconnect battery and power supplies, wait shortly, 
then reconnect 
Device defective 
Return to manufacturer for repair 
Err.1 
Measured value above allowable range 
Check: pressure not within sensor range?  
-> measuring value to high! 
Sensor defective 
Return to manufacturer for repair 
Err.2 
Measured value below allowable range 
Check: pressure not within sensor range?  
-> measuring value to low! 
Sensor defective 
Return to manufacturer for repair 
Err.3 
Display range overflow 
Check: value above 19999 -> to high to be displayed 
Err.4 
Display range underflow 
Check: value below -19999 (Tara?) -> to low 
Er.11 
Value could not be calculated 
Choose different unit 
Calculation overflow happened 
Choose different unit 
Err.7 
System error 
Return to manufacturer for repair 
– – – – 
Sensor not present / recognised 
reconnect sensor,  
during logging: stop the logger and restart it 
could not calculate value 
suitable sensor / unit combination necessary 
If 
“bAt” is flashing the battery will be exhausted soon. Further measurements are possible for short time. If 
“bAt” is displayed continuously the battery is ultimately exhausted and has to be replaced. Further 
measurements aren‟t possible any more.  
17 Reshipment und Disposal 
17.1  Reshipment 
 
_DANGER 
  
All devices returned to the manufacturer have to be free of any residual of measuring media 
and/or other hazardous substances. Measuring residuals at housing or sensor may be a risk for 
persons or environment 
 
 
 
 
Use a adequate transport package for reshipment, especially for fully functional devices. Please 
make sure that the device is protected in the package by enough packing materials.  
 
17.2  Disposal instructions 
 
Batteries must not be disposed in the regular domestic waste but at the designated collecting points.  
The device must not be disposed in the unsorted municipal waste! Send the device directly to us 
(sufficiently stamped), if it should be disposed. We will dispose the device appropriate and environmentally 