Greisinger GMH5155 Digital Manometer 603930 ユーザーズマニュアル

製品コード
603930
ページ / 44
H82.0.03.6C-08 
 Operating Manual GMH 5155 
page 21 of 22 
 
 
_____________________________________________________ 
_____________________________________________________________________________ 
 
18 Specifications 
Display range 
maximal -19999...19999 digit, depends on used sensor 
Number of channels 
Suitable probes 
GMSD 
… - K51 and MSD … sensors with connection cable MSD-K51 
Available ranges/resolutions  
ranging from  -
1.999 … 2.500 mbar / 0.001 mbar  to  0 … 1000 bar / 1 bar 
Connections 
sensor 
two 7-pole bayonet connectors 
output /  
ext. supply 
4-pole connector for serial interface and supply (USB Adapter USB 5100) 
analog output 0-1V, adjustable 
Display units 
depends on meas. range selection (depends on connected sensor): 
mbar, bar, Pa, kPa, MPa, mmHg, inHg, PSI, mH
2
User unit 
freely adjustable 
Measuring frequency 
4 / s  or 1000 / s 
Accuracy 
± 0.1 % FS ± 1 digit 
Working conditions 
-25 ... 50 °C; 0 ... 95 %RH (non condensing) 
Storage conditions 
-25 ... 70 °C 
Display 
4 ½ digit 7-segment display, illuminated (white) 
Correction function 
offset / slope (via menu),  
if offset/slope value not equal zero: "corr" is displayed 
Leakage test function 
Pressure change rate display, configurable time base 
Air velocity / volume flow  
measuring with pitot tube 
Real-time clock 
 
integrated clock with date and year 
Data logger 
cyclic 
8000 data sets 
manual 
1000 data sets (with measuring point input,  
40 adjustable measuring point text numbers) 
Alarm 
channels 
3 (sensor 1, sensor 2, difference) with individual alarm boundaries 
types 
buzzer/visual/interface 
Additional functions 
min- / max- / hold- function 
Averaging 
 
adjustable
, 1 … 120 seconds 
Housing 
protection class  IP65 and IP67 
 
dimensions 
L*W*H [mm] 
160 * 86 * 37 incl. protective covering, 
approx. 250 g incl. batteries and protective covering 
Power supply 
 
2*AAA-batteries (included in delivery)  
current 
consumption 
2.0 mA (at Out = Off, equivalent to 500 h),  
illumination ~10mA (turns off automatically) 
Battery-change indicator 
automatically if battery exhausted "bAt", warning "bAt" flashing 
Auto-off-function 
Device will be automatically switched off if no key is pressed/no interface 
communication takes place for the time of the power-off delay.  
The power-off delay can be set to values between 1and 120 min.; it can be 
completely deactivated. 
EMC 
The device corresponds to the essential protection ratings established in the 
Regulations of the Council for the Approximation of Legislation for the member 
countries regarding electromagnetic compatibility (2004/108/EG) 
Additional fault: <1%