Operating Manual GMH 2710 / GMH 2710-K
page 5 of 8
_____________________________________________________
_____________________________________________________________________________
6 Basics
Probe Precision/Device Precision
The device has a very good system accuracy (please refer to technical data). This is due to the calibration of
the probe acc. To the connected instrument plus the use of precise measurement electronics and premium
Pt1000 sensor elements
The device has a very good system accuracy (please refer to technical data). This is due to the calibration of
the probe acc. To the connected instrument plus the use of precise measurement electronics and premium
Pt1000 sensor elements
Heat loss caused by probe construction:
Especially when measuring temperatures which deviate very much from the ambient temperature, measuring
errors often occur if the heat loss caused by the probe is not considered. When measuring fluids therefore
the probe should be emerged sufficiently deep (at least 5 times of diameter, i.e.>15mm) and be stirred
continuously. When measuring gases the probe should also emerge as deep as possible in the gas to be
measured (e.g. when measuring in channel/pipes) and the gas should flow around the probe at sufficient
flow.
Especially when measuring temperatures which deviate very much from the ambient temperature, measuring
errors often occur if the heat loss caused by the probe is not considered. When measuring fluids therefore
the probe should be emerged sufficiently deep (at least 5 times of diameter, i.e.>15mm) and be stirred
continuously. When measuring gases the probe should also emerge as deep as possible in the gas to be
measured (e.g. when measuring in channel/pipes) and the gas should flow around the probe at sufficient
flow.
Allowable Temperature Range Of Probes
For the two variants there are different application areas:
For the two variants there are different application areas:
GMH 2710: (black plastic handle)
The allowable temperature range of -200 to 200 °C only is valid for the probe tube, the (plastic) handle must
not be heated above 80°C, the cable utmost 105°C! Therefore only front part of the stainless steel tube is
allowed to be heated above 80°C!
GMH 2710-K (white Teflon handle)
Both cable and handle are capable of -200 bis +250°C!
The allowable temperature range of -200 to 200 °C only is valid for the probe tube, the (plastic) handle must
not be heated above 80°C, the cable utmost 105°C! Therefore only front part of the stainless steel tube is
allowed to be heated above 80°C!
GMH 2710-K (white Teflon handle)
Both cable and handle are capable of -200 bis +250°C!
7 Configuration of the Device
Follow these instructions to configure the functions of the device:
Switch the device on.
Press
and
together until the first parameter
“UNIT” is displayed.
Set parameter with
= up or
= down.
Jump to the next parameter by pressing
.
Parameter Value
Information
button
buttons
UNIT
Display unit factory setting: °C
°C
Measuring value displayed in °Celsius
°F
Measuring value displayed in °Fahrenheit
AUTO
HLD
Auto Hold-Function factory setting:: oFF
ON
Auto Hold activated: automatic holding of the measured value, as soon as it is
stabilized.
stabilized.
OFF
Auto Hold deactivated: Value is frozen on keypress (hold)
P.OFF
Auto Power-Off (turn-off delay) factory setting: 20 min.
1 … 120
Auto Power-Off (turn-off delay) in minutes. If no key is pressed for the time
adjusted here, the device is automatically switched off (adjustable 1 … 120 min)
adjusted here, the device is automatically switched off (adjustable 1 … 120 min)
OF
Auto power-off is deactivated (continuous operation)
INIT
Restore factory settings
NO
Parameters are not changed to factory settings.
YES
ATTENTION: All parameter are changed to factory settings.
Press again to store changed settings, the device restarts (segment test).
NOTE:
If there is no key pressed within the menu mode within 2 minutes, the configuration will be
cancelled, the entered settings are lost!
cancelled, the entered settings are lost!