8  Adjustment 
The instrument can be adjusted, assuming that: Reliable references are available, such as ice-
water regulated precision water baths or similar. 

Displayed value °C =(measured value °C - OFFS) * (1 + slope correction / 100 )        
Displayed value °F =(measured value °F - 32 °F - OFFS) * (1 + slope correction / 100 )       
 
Follow these instructions to adjust the device: 
  Switch the device on. 
  Press 
 and 
 together until the first paramete
r “0FFS” is displayed.  
  Set parameter with 
 = up or 
= down  
  Jump to the next parameter by pressing 
 
Parameter  Value 
Information 
button  
buttons  
 
OFFSET correction     
factory setting: oFF = 0.0°C
 
OFF
 
No offset correction 
-2.5  2.5 °C  
or 
-4.54.5 °F
 
Value of offset correction 
Slope correction 
factory setting: oFF= 0%
 
OFF
 
No slope correction  
-5.00  5.00 
Value of slope correction in % 
 
Press          again to store changed settings, the device restarts (segment test). 
NOTE: 
If there is no key pressed within the menu mode within 2 minutes, the configuration will 
be cancelled, the entered settings are lost! 
9  Replacing Batteries 
Before changing batteries, please read the following instruction and follow it step by step.
  
Not following the instruction may cause harm to the instrument or the protection against ingress of water and 
dust may be lost!
  
Avoid unnecessary opening of the instrument! 
1.  Open the 3 Phillips screws at the backside of the instrument.  
2.  Lay down the still closed instrument, so that the display side points upwards.  
The lower half of the housing incl. The electronics should be kept lying down during battery change.  
This avoids loss of the sealing rings of the screw holes. 
3.  Lift upper half of housing. Keep an eye on the three function keys, to be sure not to damage them. 
4.  Change carefully the two batteries (Type: AAA). 
5.  Close the housing, taking care that it is positioned correctly, otherwise the sealing may be damaged. 
Afterwards press the two halves together, lay the instrument with display pointing downwards and 
screw it together again, beginning with the single lower screw.