8.3  Example applications
Overview of the pins
Rem-SB
Remote
DGnd
AGnd
RSEL
PSEL
CSEL
VSEL
VRef
R-active
R-Range
Trigger In
VMON
CMON
OVP/OT
Figure 16
Master-Slave operation, simulated
A true Master-Slave is not possible because the analogue 
interface  does  not  provide  set  value  outputs.  But  the 
monitor outputs CMON or, in some cases, even VMON 
of the master can be used to control at least one of the 
four set value inputs of one or multiple slave loads. 
AGnd
CMON
DGnd PSEL
Master
Slave
Figure 17
Since there is no power monitor output, the PSEL input(s) 
can‘t be controlled by the master. But it could be directly 
tied to VREF output or via a potentiometer, in order to 
adjust the power between 0% and 100%. The inputs 
Re-
mote and R-active have to be tied to GND at the slave(s) 
in order to activate the external control.
Input off
Figure 18 shows the wiring of the analogue interface for 
remotely switching the input off. This feature can be used 
anytime and does not require the activation of external 
control by pin 
Remote. It can be combined with other 
applications  and  can  be  realised  by  various  contacts, 
like transistors, relays, switches etc. Opening the contact 
again will either switch the output on if it was on before 
switching off (remote control) or enable switching it on 
again manually on the front panel.
DGND
REM-SB
DGND
REMOTE
Figure 18
Switching to external control
Switching to external control is only required if the device 
is going to be controlled by external analogue signals. 
If using the simulated master-slave operation, only the 
slave(s)  have  to  be  switched  to  external  control. The 
switch-over is realised with either a relay or a switch etc.
Important! All set values have to be given, except the 
resistance set value which is only required if resistance 
control is activated. Set values can also be bridged to 
pin VREF for 100%.
DGND
REM-SB
DGND
REMOTE
Figure 19
External control with current and power
The  example  in  figure  20  shows  potentiometers,  one 
each for the set value of power and current, which are 
referenced to VREF (10V) and AGND. With this you can 
arbitrarily  adjust  current  and  power  between  0%  and 
100%. Voltage set value VSEL and signal R-active tied 
to 0V here for correct operation.
VREF
AGND
CSEL
10k
DGND
PSEL
Figure 20
Remote control