3.9.4  Manual operation
3.9.4.1  Function selection and control
Via the touchscreen one of the functions described above can be called up, 
configured and controlled. Selection and configuration are only possible when 
the input is switched off.
 
► How to select a function and adjust parameters
1. 
Tap the touch area 
 on the main screen. If the menu does not appear it is because the DC input 
is still switched on or the touch area is locked due to the device being possibly in remote control.
2. 
In the menu overview tap on the touch area 
 and then on the desired function.
 Note: this touch area is locked in master-slave mode or in R mode (adjustable resistance).
3. 
Depending on the choice of function there follows a request to which value the function generator is going 
to be applied: 
 or 
.
4. 
Adjust the parameters as you desire, like offset, amplitude and frequency for a sine wave, for example.
For the AC part of a function and if the difference between start and end value of amplitude 
or frequency is too low (min. increment of 0.58%), depending on the time that is defined for 
one function run, the function generator will not accept the settings and pop up an error.
5. 
Do  not forget to adjust the overall limits of voltage, current and power, which you can access with touch 
area 
When entering function generator mode, those global limits are reset to safe values, which 
might prevent the function from working at all. For example, if you apply the selected 
function to the input current, then the overall current limit should not interfere and should 
at least be as high as offset + amplitude.
Setting the various functions is described below. After setting it up, the function can be loaded.
 
► How to load a function
1. 
After setting the values for the required signal generation, tap on the 
touch area 
The device will then load the data into the internal controller and changes the 
display. Shortly afterwards the static values are set (power and voltage or 
current), the DC input is switched on and the touch area 
 enabled. 
Only then can the function be started.
The static values are applied to the DC input immediately after loading the function, because it 
switches the DC input on automatically in order to settle the start situation. These static values 
represent start and end values for the progress of the function, so that the function does not 
need to start from 0. Only exception: when applying any function to the current (I), there is no 
adjustable static current value, so the function would always start from 0 A.