Leica CM1950 ユーザーズマニュアル
22
Operating Manual V 1.1 – 05/2007
•
Insert the pin (1) of the handwheel shaft into
the hole (2) of the handwheel.
the hole (2) of the handwheel.
•
Tighten the screw (4) using the No. 6 Allen key.
•
Place the protective cap on the screw (4).
To dismount, proceed in reverse order.
1
2
4
6.3
Assembling the Handwheel
Fig. 14
Transport with a fork lift
•
The instrument can be transported with a fork lift.
To ensure a safe transport with a fork lift 3 people are required:
one operating the fork lift, and the other 2 holding the instrument
on either side to prevent it from sliding down.
one operating the fork lift, and the other 2 holding the instrument
on either side to prevent it from sliding down.
•
At the installation site, unscrew the adjustable feet (1) with the open-
end wrench No. 13 (see Fig. 12). This is absolutely necessary for a stable
stand.
end wrench No. 13 (see Fig. 12). This is absolutely necessary for a stable
stand.
Fig. 13
Total width with handwheel
6.
Installation
Rotate the handwheel only if the
refrigeration system is on and the
cryochamber is cold.
refrigeration system is on and the
cryochamber is cold.