Leica CM1950 Manuel D’Utilisation

Page de 74
22
Operating Manual V 1.1 – 05/2007
Insert the pin (1) of the handwheel shaft into
the hole (2) of the handwheel.
Tighten the screw (4) using the No. 6 Allen key.
Place the protective cap on the screw (4).
To dismount, proceed in reverse order.
1
2
4
6.3
Assembling the Handwheel
Fig. 14
Transport with a fork lift
The instrument can be transported with a fork lift.
To ensure a safe transport with a fork lift 3 people are required:
one operating the fork lift, and the other 2 holding the instrument
on either side to prevent it from sliding down.
At the installation site, unscrew the adjustable feet (1) with the open-
end wrench No. 13 (see Fig. 12). This is absolutely necessary for a stable
stand.
Fig. 13
Total width with handwheel
6.
Installation
Rotate the handwheel only if the
refrigeration system is on and the
cryochamber is cold.